Clinical validation of a commercial LAMP Test for ruling out Malaria in returning travelers: a prospective diagnostic trial

Thumbnail
Fecha
2018
Ver/
Clinical validation of a commercial LAMP Test for ruling out Malaria in returning travelers: a prospective diagnostic trial.pdf (257.0Kb)
Metadatos
Mostrar el registro completo del ítem
URI
http://patua.iec.gov.br//handle/iec/3632
Autor
Cheaveau, James
Nguyen, Hong
Chow, Barbara
Marasinghe, Dewdunee
Mohon, Abu Naser
Yuan, Hong
Viana, Giselle Maria Rachid
Van Schalkwyk, Donelly
Church, Deirdre
Chan, Wilson
Pillai, Dylan R
Resumen
The mainstay of malaria diagnosis relies on rapid diagnostic tests (RDTs) and microscopy, both of which lack analytical sensitivity. This leads to repeat testing to rule out malaria. A prospective diagnostic trial of the Meridian illumigene Malaria assay (loop-mediated isothermal amplification [LAMP]) was conducted comparing it with reference microscopy and RDTs (BinaxNOW Malaria) in returning travelers between June 2017 and January 2018. Returning travelers with signs and symptoms of malaria were enrolled in the study. RDTs, microscopy, and LAMP assays were performed simultaneously. A total of 298 patients (50.7% male; mean age, 32.5 years) were enrolled, most visiting friends and relatives (43.3%), presenting with fever (88.9%), not taking prophylaxis (82.9%), and treated as outpatients (84.1%). In the prospective arm (n = 348), LAMP had a sensitivity of 98.1% (95% confidence interval [CI], 90.0%-100%) and a specificity of 97.6% (95% CI, 95.2%-99.1%) vs microscopy. After discrepant resolution with real-time polymerase chain reaction, LAMP had a sensitivity of 100% (95% CI, 93.7%-100%) and a specificity of 100% (95% CI, 98.7%-100%) vs microscopy. After discrepant resolution, RDTs had a sensitivity of 83.3% (95% CI, 58.6%-96.4%) and a specificity of 96.2% (95% CI, 93.2%-98.1%) vs microscopy. When including retrospective specimens (n = 377), LAMP had a sensitivity of 98.8% (95% CI, 93.2%-100%) and a specificity of 97.6% (95% CI, 95.2%-99.1%) vs microscopy, and after discrepant resolution of this set, LAMP had a sensitivity of 100% (95% CI, 95.8%-100%) and a specificity of 100% (95% CI, 98.7%-100%). A cost-benefit analysis of reagents and labor suggests savings of up to USD$13 per specimen using a novel algorithm with LAMP screening.
Referencia
CHEAVEAU, James et al. Clinical validation of a commercial LAMP Test for ruling out Malaria in returning travelers: a prospective diagnostic trial. Open Forum Infectious Diseases, v. 5, n. 11, p. 1-9, Oct. 2018.
DeCs
Malária / diagnóstico
Malária / transmissão
Saúde do Viajante
Reação em Cadeia da Polimerase
Testes Sorológicos / métodos
Sensibilidade e Especificidade
Estudos de Casos e Controles
Derecho de autor
Acesso Aberto
Colecciones
  • SAPAR - Artigos Científicos

Instituto Evandro Chagas - SVS - MS - 2007-2018 Rodovia BR316 km 7 sn - Levilandia - 67030-000 - Ananindeua - Para - Brasil.
Licença Creative CommonsEste trabalho está licenciado com uma Licença Creative Commons - Atribuição-NãoComercial 4.0 Internacional
Tel: (55 91) 3214-2191
Email: biblioteca@iec.gov.br / clariceneta@iec.gov.br
 

 

Listar

Todo DSpaceComunidades & ColeccionesPor fecha de publicaciónAutoresTítulosMateriasEsta colecciónPor fecha de publicaciónAutoresTítulosMaterias

Mi cuenta

AccederRegistro

Instituto Evandro Chagas - SVS - MS - 2007-2018 Rodovia BR316 km 7 sn - Levilandia - 67030-000 - Ananindeua - Para - Brasil.
Licença Creative CommonsEste trabalho está licenciado com uma Licença Creative Commons - Atribuição-NãoComercial 4.0 Internacional
Tel: (55 91) 3214-2191
Email: biblioteca@iec.gov.br / clariceneta@iec.gov.br